• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 ACETABULAR SCREW 6.5MM X 20MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 ACETABULAR SCREW 6.5MM X 20MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/04/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: cat# ep-200160 lot# 507990 act artic e1 hip brg 28x54mm.Cat# 110024467 lot# 695130 g7 dual mobility liner 54mm i.Cat# 650-1055 lot# 3166135 cer bioloxd option hd 28mm.Cat# 650-1064 lot# 3162350 cer option type 1 tpr sleve -6.G2: foreign ¿ australia.The customer has indicated that the product will not be returned to zimmer biomet for investigation as is location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that approximately three days post-implantation, the patient underwent a left hip revision due to cup loosening.The cup, screws, liner, bearing, and head were revised.The stem was retained.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: g3; h2; h3; h6; h10.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were provided and review of the available records identified the following: fracture of posterior and superior wall with cup loosening.Radiographs identified the following: loose acetabular cup with displacement and vertical cup orientation.A pelvic fracture is not identified in the x-ray views, but cannot be excluded given the cup displacement.A definitive root cause cannot be determined.The patient had previous trauma requiring pelvic surgery.It is unknown if this past history contributed to the reported event.The complaint is confirmed due to the x-ray findings and medical record review.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 ACETABULAR SCREW 6.5MM X 20MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19386864
Report Number0001825034-2024-01350
Device Sequence Number11767110
Product Code PBI
UDI-Device Identifier00880304527454
UDI-Public(01)00880304527454(17)330701(10)7557579
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number010000997
Device Lot Number7557579
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2024
Supplement Date Received by Manufacturer09/19/2024
Initial Report FDA Received Date05/23/2024
Supplement Report FDA Received Date09/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age51 YR
Patient SexFemale
Patient Weight85 KG
-
-