MAUDE Adverse Event Report: LIVANOVA USA, INC. PULSE GEN MODEL 1000-D; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. PULSE GEN MODEL 1000-D; GENERATOR Back to Search Results
Model Number 1000-D
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Post Operative Wound Infection (2446)
Date of Event 04/28/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H3.Device evaluated by mfr?; code 81 - device evaluation is not necessary because the return and evaluation would add no value to the investigation.Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event or Problem Description
It was reported that the patient had their generator explanted with a wound revision/washout due to infection.Device history records were reviewed for the generator and lead.The generator and lead passed all specifications prior to distribution.The generator and lead were hp sterilized.No other relevant information has been received to date.
 
Event or Problem Description
Update was received from the physician.The infection began on (b)(6) 2024 and the cause of the infection is unknown, but possibly due to manipulation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 1000-D
Common Device Name
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key19386920
Report Number1644487-2024-00615
Device Sequence Number7072114
Product Code LYJ
UDI-Device Identifier05425025750528
UDI-Public05425025750528
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P970003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/23/2025
Device Model Number1000-D
Device Lot Number7478
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2024
Supplement Date Received by Manufacturer05/28/2024
Initial Report FDA Received Date05/23/2024
Supplement Report FDA Received Date06/21/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient SexFemale
-
-