ORIDION MEDICAL MICROSTREAM ADVANCE; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE Back to Search Results
Model Number MVIIH
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/22/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during use, the unit had a decrease in numerical measurement, it was coming out below the normal range and it was falling compared to other devices.There was no reported patient outcome.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted dents along the tubing.The tubing was looked at under a microscope.Some of the dents were found to be little pin holes through the tubing.There were also moisture and debris rings throughout the tubing and in the intubation adaptor that were narrowing the diameter of the tubing and the opening in the adaptor.No other faults were found with the filterline.It was reported that the unit had a decrease in numerical measurement, it was coming out below the normal range and it was falling compared to other devices.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROSTREAM ADVANCE
Common Device Name
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
ORIDION MEDICAL
7 hamarpeh st.
jerusalem 91235
IS  91235
Manufacturer (Section G)
ORIDION MEDICAL
7 hamarpeh st.
jerusalem 91235
IS   91235
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19387880
Report Number8044004-2024-00011
Device Sequence Number19838453
Product Code CCK
UDI-Device Identifier20884521761985
UDI-Public20884521761985
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K181467
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMVIIH
Device Catalogue NumberMVIIH
Device Lot NumberC230417501
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2024
Supplement Date Received by Manufacturer02/20/2025
Initial Report FDA Received Date05/23/2024
Supplement Report FDA Received Date02/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-