U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. AVL ANGLED REAMER SLEEVE ASSEMBLY, 15; ORTHOPEDIC MANUAL SURG INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. AVL ANGLED REAMER SLEEVE ASSEMBLY, 15; ORTHOPEDIC MANUAL SURG INSTR Back to Search Results
Model Number AVL ANGLED REAMER SLEEVE ASSEMBLY, 15
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
On 07/01/2022, it was reported by a sales representative via sems that during an arthroplasty procedure a ar-9676 reamer, and a ar-9297-15 reamer sleeve cold welded together.Also a ar-9239ag drill was found to have a piece of metal chipped.There was no patient effect reported.No additional information reported.No additional information provided.Additional information was provided on 07/07/2022, it was reported that the ar ¿ 9239 ag did not chip inside the body, after the drilling was complete the tech put the drill on the table and it fell off onto the floor and this caused the chip in the drill bits.The case was completed successfully.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to user error of the device due to prying/leveraging the device during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVL ANGLED REAMER SLEEVE ASSEMBLY, 15
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19387894
Report Number1220246-2024-03797
Device Sequence Number11767252
Product Code LXH
UDI-Device Identifier00888867350038
UDI-Public00888867350038
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAVL ANGLED REAMER SLEEVE ASSEMBLY, 15
Device Catalogue NumberAR-9297-15
Device Lot Number052043
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2022
Initial Report FDA Received Date05/23/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/15/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-