| Model Number |
71926-01 |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Codes |
Loss of consciousness (2418); Convulsion/Seizure (4406)
|
| Date of Event |
04/30/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.Sensor manufacturing date: 07-feb-24.Sensor expiration date: 30-sep-24.The most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.The customer reported signal loss.Adc attempted to be replicated using similar configurations of iphone 12 (ios 17.1.2, 2.7.5.4061).The reported issue was unable to be replicated and the system functioned as intended.There were no issues identified with the customer's reported application during replication that would have led to the reported issue.The date of event is unknown.The date entered in section b3 is the date abbott diabetes care became aware of the event.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
An alarm issue was reported with the abbott diabetes care (adc) device in use with the iphone 13 with ios operating system version 17.1.1.The customer experienced a signal loss and was unable to obtain readings and receive glucose alarms.As a result, the customer was not alerted of changes in glucose level, experienced seizures and a loss of consciousness, and was able to self-treat (unspecified).No third-party treatment was reported.There was no report of death or permanent impairment associated with this event.
|
| |
|
Event or Problem Description
|
|
An alarm issue was reported with the abbott diabetes care (adc) device in use with the iphone 13 with ios operating system version 17.1.1.The customer experienced a signal loss and was unable to obtain readings and receive glucose alarms.As a result, the customer was not alerted of changes in glucose level, experienced seizures and a loss of consciousness, and was able to self-treat (unspecified).No third-party treatment was reported.There was no report of death or permanent impairment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.The most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time no product has been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.The customer reported signal loss.Adc attempted to be replicated using similar configurations.The reported issue was unable to be replicated and the system functioned as intended.There were no issues identified with the customer's reported application during replication that would have led to the reported issue.If the product is returned, a physical investigation will be performed and a follow-up report submitted.This serves as a correction report.Section h11 (additional mfg narrative) was incorrectly submitted in the previous report.Correction has been made.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|