| Model Number |
K172 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Off-Label Use (1494)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/30/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker recorded noise over sensing during a non-conditional magnetic resonance imaging (mri) procedure.Ventricular pacing marched through the observed noise at the mri procedure.The pacemaker remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker recorded noise over sensing during a non-conditional magnetic resonance imaging (mri) procedure.Ventricular pacing marched through the observed noise at the mri procedure.The pacemaker has been explanted due to normal battery depletion (nbd) and returned for analysis without additional allegations.No adverse patient effects were reported.
|
| |
|