Catalog Number 33680013
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 03/13/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event could be confirmed based on available medical record and medical expert assessment.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Microbiologist reviewed the sterilization procedures, environmental monitoring, bioburden data and the dhr and noted: the subject device was packaged according to established design and process specifications, and was sterilized according to procedure.No deviation for a non-conformance could be found.Upon further investigation of the ct scans by healthcare professionals the following was observed ¿there is some radiolucence and an anterior defect in the tibial bone.Together with the clinical information given, loosening can be confirmed.There is also infection possible in this case.Given the time between the implantation and the need for a revision the infection is patient related and not related to the implant or the implantation.¿ based on investigation, the root cause was attributed to a patient related issue.The failure was detected due to some radiolucency and an anterior defect in the tibial bone.Hence, loosening is confirmed.Also, according to the hcp, an infection is also possible given the time between the implantation and need for a revision, the infection is patient related and not related to the implant, but this cannot be confirmed.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery due to talar bone loss.The main reason for revision is aseptic loosening.Unsure if there's infection there.Cultures has been negative but has elevated esr and crp.As a result, will consider open culture 1st and if negative, will go ahead with revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY ADAPTIS TIB SZ3LNG INFINITY ADAPTIS
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19388791
Report Number3010667733-2024-00308
Device Sequence Number12501397
Product Code HSN
UDI-Device Identifier00889797069687
UDI-Public00889797069687
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K181557
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number33680013
Device Lot Number1740309
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/29/2024
Initial Report FDA Received Date05/23/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-