• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES PR FOGARTY ARTERIAL EMBOLECTOMY CATHETER; CATHETER, EMBOLECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES PR FOGARTY ARTERIAL EMBOLECTOMY CATHETER; CATHETER, EMBOLECTOMY Back to Search Results
Model Number 120804FP
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported via medwatch (b)(4) the tip of a fogarty catheter looked chipped and slightly frayed.The balloon would not inflate and saline leaked from multiple holes on all catheters.Procedure was completed with another catheter.There were no patient harm.A product evaluation of the returned 120804f was returned.The reported event of balloon issue was confirmed.Balloon latex appeared deteriorated.Multiple cracks and a tear were evident on balloon latex.Leakage was observed through the tear on the balloon.Both balloon windings were intact.Balloon latex was released from proximal winding to check balloon edges at the tear.The edges did not appear to match at the tear.No other visible damage was observed from catheter body.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Additional Manufacturer Narrative
Corrections to h10 was made.Previous submission referenced (b)(4), but was not included in the related report number field.
 
Additional Manufacturer Narrative
An engineering evaluation was initiated to assess for any manufacturing related processes which could be correlated to the complaint.Based on the available information, there is no evidence that supports or confirms the failure mode is associated to a manufacturing or design defect.Capa was previously initiated and used as reference in the evaluation as it addresses the "balloon - latex deterioration" failure for the embolectomy models with a vascupouch packaging.The capa investigation concluded that the root cause is exposure of latex to ozone, which can happen when the pouch packaged device is stored in rooms with high energy ionizing radiation sources that could generate ozone e.G.Fluoroscopy machines, x ray machines, uv lights, hvac sanitation equipment, etc.The storage conditions for these catheters are specified in the ifu.The capa resulted into a voluntary field correction action that instructed customers to return any product which has been stored in the same room as high energy ionizing radiation sources which emit ozone.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOGARTY ARTERIAL EMBOLECTOMY CATHETER
Common Device Name
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco
Manufacturer Contact
katheryn richards
1 edwards way
irvine, CA 92614
9492017706
MDR Report Key19389076
Report Number2015691-2024-03869
Device Sequence Number6786270
Product Code DXE
UDI-Device Identifier00690103205213
UDI-Public(01)00690103205213(17)240718(11)220419(10)64311190
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date07/18/2024
Device Model Number120804FP
Device Lot Number64311190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/23/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2024
Supplement Date Received by Manufacturer05/30/2024
06/04/2024
Initial Report FDA Received Date05/23/2024
Supplement Report FDA Received Date05/30/2024
06/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-