This device is classified as import for export, therefore 510k is not applicable.Model fb-18v is available in the usa with a 510k number k951199.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ocular leak.Based on the result, we concluded that it was caused due to the physical damage applied on the ocular.In addition, our technician confirmed that the segment broken, the insertion flexible tube buckled, the angle wire play, the body cover grip scratched, the distal body dirty, the objective lens dirty, the lcb distal cover glass dirty, and the root brace rubber (lg control body) flared; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|