This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023. arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint confirmed.One unpackaged ar-8656-02, batch 1392111 chondral pick was received for investigation.Visual inspection identified that the distal end of the shaft of the chondral pick had a significant warp.The thickness of the tip of the pick as well as of the base of the shaft was assessed, and each dimension was found to meet design specifications.Additionally, gouges were present along the outer diameter of the distal end of the shaft.The bone quality experienced during the procedure was not recorded.The most likely cause for the reported failure can be attributed to misuse/mishandling due to damage to the device.
|