Model Number 40495
Medical Device Problem Codes Entrapment of Device (1212); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/15/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the guidewire was stuck in the catheter.A 2x40x90 sterling balloon catheter was advanced for dilation along with a 200cm, 8cm v18 control guidewire.During the procedure, the balloon was inflated and deflated successfully.Upon removal of the balloon, it was not able to be removed from the wire.The physician tried to pull it back gently over the wire, but it did not resolve the issue.Subsequently, both the balloon and the guidewire were taken out together from the patient, and successfully managed to remove the wire from the balloon shaft.The procedure was completed with another v-18 wire.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the device was returned for analysis.The device was returned for the analysis, and it is possible to see that the guidewire was bent distal tip and middle section.Also, the original pouch was returned, and it was observed that the pouch information matches with the complaint information.Based on the evidence, the as reported code body entrapment of device or device component can be confirmed due the device condition.Boston scientific manufacturing processes include extensive inspections to ensure that all finished devices meet specifications.
 
Event or Problem Description
It was reported that the guidewire was stuck in the catheter.A 2x40x90 sterling balloon catheter was advanced for dilation along with a 200cm, 8cm v18 control guidewire.During the procedure, the balloon was inflated and deflated successfully.Upon removal of the balloon, it was not able to be removed from the wire.The physician tried to pull it back gently over the wire, but it did not resolve the issue.Subsequently, both the balloon and the guidewire were taken out together from the patient, and successfully managed to remove the wire from the balloon shaft.The procedure was completed with another v-18 wire.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-18 CONTROL WIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key19391294
Report Number2124215-2024-30085
Device Sequence Number7049377
Product Code DQX
UDI-Device Identifier08714729070191
UDI-Public08714729070191
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K033742
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/26/2025
Device Model Number40495
Device Catalogue Number40495
Device Lot Number0032738670
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2024
Supplement Date Received by Manufacturer06/24/2024
Initial Report FDA Received Date05/24/2024
Supplement Report FDA Received Date07/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BALLOON: STERLING; BALLOON: STERLING
Patient SexUnknown
-
-