| Model Number |
40495 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the guidewire was stuck in the catheter.A 2x40x90 sterling balloon catheter was advanced for dilation along with a 200cm, 8cm v18 control guidewire.During the procedure, the balloon was inflated and deflated successfully.Upon removal of the balloon, it was not able to be removed from the wire.The physician tried to pull it back gently over the wire, but it did not resolve the issue.Subsequently, both the balloon and the guidewire were taken out together from the patient, and successfully managed to remove the wire from the balloon shaft.The procedure was completed with another v-18 wire.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was returned for analysis.The device was returned for the analysis, and it is possible to see that the guidewire was bent distal tip and middle section.Also, the original pouch was returned, and it was observed that the pouch information matches with the complaint information.Based on the evidence, the as reported code body entrapment of device or device component can be confirmed due the device condition.Boston scientific manufacturing processes include extensive inspections to ensure that all finished devices meet specifications.
|
| |
|
Event or Problem Description
|
|
It was reported that the guidewire was stuck in the catheter.A 2x40x90 sterling balloon catheter was advanced for dilation along with a 200cm, 8cm v18 control guidewire.During the procedure, the balloon was inflated and deflated successfully.Upon removal of the balloon, it was not able to be removed from the wire.The physician tried to pull it back gently over the wire, but it did not resolve the issue.Subsequently, both the balloon and the guidewire were taken out together from the patient, and successfully managed to remove the wire from the balloon shaft.The procedure was completed with another v-18 wire.There were no patient complications reported.
|
| |
|