| Model Number |
26167FNS |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
05/16/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.Internal karl storz reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior promontofixation with placement of tapes via laparoscopy, one of the jaws of the needle holder broke during the introduction of the instrument into the abdomen.After searching the environment and surgical exploration, the broken jaw could not be found.Presence in the patient's abdomen of a 2x8.5mm piece of metal found during an unprepared abdominal x-ray on (b)(6) 2024.Actions taken in the care facility to manage the patient: inform the patient of the risks.Due to the fragement remained in situ a report is required.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was not returned for investigation.Investigation was carried out on 2024-07-23.Based on the customer's description of the fault, a mechanical overload caused by the user can be assumed.One possible cause of the breakage could be, for example, that the needle was gripped outside of the gripping area and the resulting forces caused the jaw to break.On the other hand, the expected number of uses and reprocessing could have weakened the material.As the article was not sent in for investigation, more detailed results cannot be provided.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Search Alerts/Recalls
|