Model Number 1013465-060
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the peroneal artery.The 3x60mm armada 18 balloon dilatation catheter (bdc) was advanced to the target lesion and the balloon was attempted to be inflated twice, however, the balloon was noted to be leaking contrast.Therefore, the bdc was removed from the patient.There was no adverse patient effect and no clinically significant delay reported in the procedure.After removal, outside the patient the balloon was injected with contrast again and a hole was noted in the balloon.Another armada was used to complete the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Visual and functional analysis was performed on the returned device.The reported balloon rupture was confirmed.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the reported information and results of the complaint investigation there is no indication the reported balloon rupture is related to a potential product quality issue.Balloon material ruptures can be affected by numerous factors including, but not limited to, balloon damage during processing of the balloon material, materials, inflation technique, and interactions with other devices or lesion calcification.There was no leak noted during preparation or during the first inflation, which suggests a product quality issue did not contribute to the reported difficulties.In this case, it is possible that the balloon interacted with the patient¿s anatomy such that the balloon became damaged and subsequently ruptured during inflation; however, a conclusive cause cannot be determined since limited information was provided.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.D1 correction: brand name updated d2a correction: common device name updated d3 correction: name, address updated e1 - physician's first name/last name updated from (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 2807
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key19393055
Report Number2024168-2024-06266
Device Sequence Number15192323
Product Code LIT
UDI-Device Identifier08717648207877
UDI-Public(01)08717648207877(17)250331(10)1102941
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model Number1013465-060
Device Catalogue Number1013465-060
Device Lot Number1102941
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/08/2024
Supplement Date Received by Manufacturer07/01/2024
Initial Report FDA Received Date05/24/2024
Supplement Report FDA Received Date07/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-