Model Number C2TR01
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Rash (2033); Fungal Infection (2419)
Date of Event 04/19/2011
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by the patient that they had a skin rash that had become quite severe.The patient had seen a dermatologist for years for the skin rash that developed shortly after original implant.The rash was initially thought to be eczema.The patient also stated that after the most recent device change out, the rash has gotten progressively worse.The dermatologist performed a biopsy and determined the skin rash was due to the patient being allergic to yeast and not anything "in their device".The yeast infection will be treated by the physician.The crt-p has been explanted and the leads remained in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCRA CRT-P
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key19395005
Report Number9614453-2024-02054
Device Sequence Number7799573
Product Code NVZ
UDI-Device Identifier00613994753304
UDI-Public00613994753304
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P010015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/28/2012
Device Model NumberC2TR01
Device Catalogue NumberC2TR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2024
Initial Report FDA Received Date05/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/06/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4194-78 LEAD, 5076-52 LEAD, 5076-45 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age68 YR
Patient SexMale
Patient Weight89 KG
-
-