• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH CHUCK-KEYLESS F/BPL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH CHUCK-KEYLESS F/BPL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.731#S-P#530.731
Medical Device Problem Codes Mechanical Jam (2983); Complete Loss of Power (4015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.The keyless chuck device was evaluated and the reported condition that the device did not function was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device was frozen/will not move.The assignable root cause of this condition was determined to be traced to component failure due to wear.Udi - (b)(4).
 
Event or Problem Description
It was reported from india that during service and evaluation, it was determined that the keyless chuck device was frozen/will not move.It was further determined that the device failed pretest for check drill function.It was noted in the service order that the device was not working.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2024.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.Investigation summary : the actual device was returned for evaluation.During repair, an evaluation was performed; and it was determined that the device had the following failure: frozen/will not move - chuck seized; and failed pretest on the following: check drill chuck function.Therefore, the reported condition that the device was not working was confirmed.The assignable root cause was determined to be due to component wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHUCK-KEYLESS F/BPL
Common Device Name
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key19398148
Report Number8030965-2024-06677
Device Sequence Number20571728
Product Code GEY
UDI-Device Identifier10886982240163
UDI-Public10886982240163
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number530.731#S-P#530.731
Device Lot Number116L346
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2024
Supplement Date Received by Manufacturer07/05/2024
Initial Report FDA Received Date05/25/2024
Supplement Report FDA Received Date07/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/12/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-