• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SP FBRTAK RC FBRTPE BLK/BLU & STTPE BLU; NON-DEGRADABLE FIXATN FASTENER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SP FBRTAK RC FBRTPE BLK/BLU & STTPE BLU; NON-DEGRADABLE FIXATN FASTENER Back to Search Results
Model Number SP FBRTAK RC FBRTPE BLK/BLU & STTPE BLU
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
On 1/13/2023, it was reported by a sales representative via email that an ar-3652ttsp fibertak rc suture anchor would not deploy correctly and pulled out of implantation site.This was discovered during a procedure on (b)(6) 2023.Additional information requested.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to user-applied excessive force, improper bone prep; misaligned insertion; prying/leveraging the device during insertion.As no further investigation was able to be performed no change in harm was identified.This complaint will be included in trending per procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SP FBRTAK RC FBRTPE BLK/BLU & STTPE BLU
Common Device Name
NON-DEGRADABLE FIXATN FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19398621
Report Number1220246-2024-03944
Device Sequence Number12703328
Product Code MBI
UDI-Device Identifier10888867362946
UDI-Public10888867362946
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K203268
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSP FBRTAK RC FBRTPE BLK/BLU & STTPE BLU
Device Catalogue NumberAR-3652TTSP
Device Lot Number14950993
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2023
Initial Report FDA Received Date05/25/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-