| Model Number |
AVL ANGLED REAMER SLEEVE ASSEMBLY, 15 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint is not confirmed.Visual evaluation revealed that the device was not cold welded to the drive shaft, abrasion marks on the base of the angled reamer, and inside of the inner diameter of the device were found.No problems found.
|
| |
|
Event or Problem Description
|
|
On (b)(6)2023, it was reported by a sales representative via sems that an ar-9297-15 angled reamer sleeve, and an ar-9676 drive shaft cold welded together.This occurred during an augmented vaultlock on (b)(6)2023.The case was completed, with no patient effect reported.There was no additional information provided.Additional information has been requested.Additional information was provided on 05/16/2023, it was reported that sufficient reaming had been performed before the device failed.
|
| |
|