| Model Number |
CCWTT0 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Visual Impairment (2138)
|
| Date of Event |
04/17/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A non-health care professional reported that following an intraocular lens (iol) implant procedure, the patient experienced with poor vision.The lens was exchanged for another lens model 21 days following the initial procedure.Additional information was received and stated that patient was not tolerating the lens, patient issues were resolved after iol exchange and there was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The product investigation could not identify a root cause for the reported complaint.Each lens is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the multifocal company, 20.0 diopter (add power +2.2/+3.2) lens was replaced with a non-company, 19.5 diopter, monofocal lens model.The product has not been received to evaluate.Due to the exchange of a multifocal lens to a monofocal lens, this may suggest that the replacement lens model was an improved choice for the patient's vision needs.No additional information has been provided at this time.Due diligence has been performed in an attempt to obtain further information on this event.File will be reopened if new information or the sample is received.The manufacturer internal reference number is:(b)(4).
|
| |
|
Search Alerts/Recalls
|