| Catalog Number |
197520075 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Limb Fracture (4518)
|
| Date of Event |
06/07/2013
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Pc was created to capture the event.On (b)(6) 2013, there is a new transverse periprosthetic fracture of the left greater trochanter periprosthetic fracture of the left greater trochanter with mild superior displacement of the fracture fragment.Intact bilateral total hip arthroplasties with left proximal femoral cerclage wires.Doi: (b)(6) 2012.Doe: (b)(6) 2013.Left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: pc was created to capture the event.(b)(6) 2013, there is a new transverse periprosthetic fracture of the left greater trochanter periprosthetic fracture of the left greater trochanter with mild superior displacement of the fracture fragment.Intact bilateral total hip arthroplasties with left proximal femoral cerclage wires.The product was not returned to j&j medtech orthopaedics; however, photos were provided for review.See attachment (b)(4) xrays retrieved by (b)(6).The x-ray investigation revealed nothing indicative of a device nonconformance at the reclaim prx bdy cone 20x75.A manufacturing record evaluation was performed for the finished device product description: reclaim prx bdy cone 20x75 product code: 197520075 lot number: 134693 and no non-conformances or manufacturing irregularities were identified.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the reclaim prx bdy cone 20x75 would have not contributed to the reported adverse event.¿ based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device product description: reclaim prx bdy cone 20x75 product code: 197520075 lot number: 134693 and no non-conformances or manufacturing irregularities were identified.H11 additional narrative: corrected: d4 (primary udi number).
|
| |
|
Search Alerts/Recalls
|