| Catalog Number |
MN1213 |
| Medical Device Problem Codes |
Break (1069); Failure to Obtain Sample (2533)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiry date: 04/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a biopsy procedure, after firing the needle with the device, the device allegedly failed to fire a second shot and it could not collect the fragment.It was further reported that the device was opened and the needle was allegedly found to be broken inside the instrument.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during a biopsy procedure, after firing the needle with the device, the device allegedly failed to fire a second shot and it could not collect the fragment.It was further reported that the device was opened and the needle was allegedly found to be broken inside the instrument.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported failure to obtain sample and break issues could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 04/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a biopsy procedure, after firing the needle with the device, the device allegedly failed to fire a second shot and it could not collect the fragment.It was further reported that the device was opened and the needle was allegedly found to be broken inside the instrument.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: additional information was received, and the file was reassessed for reportability, and determined to be no longer reportable.Since an initial mdr was submitted, therefore, the file will remain assessed as a malfunction.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|