| Model Number |
CR2 |
| Medical Device Problem Code |
No Audible Prompt/Feedback (2282)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device would not give audio prompts when opened.In this state the device may not be able to deliver defibrillation therapy if needed.There was no patient involvement reported with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker product analysis center (pac) evaluated the customer's device and was able to duplicated the reported event.Stryker pac determined the reed switch sw5 was the cause of the reported issue.The device was archived by stryker and the customer received a replacement device.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device would not give audio prompts when opened.In this state the device may not be able to deliver defibrillation therapy if needed.There was no patient involvement reported with the event.
|
| |
|