Model Number 15
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker performed an initial evaluation of the customer's device and verified the reported issue.Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted stryker to report that their device had a completely white display.A white display can be indicative of a device lock up condition.As a result, defibrillation therapy may be unavailable, if needed.There was no report of patient use associated with the reported event.
 
Additional Manufacturer Narrative
Stryker replaced the device's user interface pcb assembly to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.Stryker further evaluated the removed user interface pcb assembly and determined that the cause of the reported issue was due to a cracked and shorted capacitor, designator c181.
 
Event or Problem Description
The customer contacted stryker to report that their device had a completely white display.A white display can be indicative of a device lock up condition.As a result, defibrillation therapy may be unavailable, if needed.There was no report of patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 20E DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key19423648
Report Number0003015876-2024-01771
Device Sequence Number20916808
Product Code MKJ
UDI-Device Identifier00883873853863
UDI-Public00883873853863
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K130454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99507-000122
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/30/2024
Supplement Date Received by Manufacturer09/18/2024
Initial Report FDA Received Date05/30/2024
Supplement Report FDA Received Date10/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/29/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-