BOSTON SCIENTIFIC CORPORATION ESSENTIO SR; IMPLANTABLE DEVICE Back to Search Results
Model Number L100
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Review of device memory confirmed that a low voltage alertcode 1003) was recorded>>.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of devices in the accolade pacemaker family that have a potential of exhibiting this behavior.This particular device is included in the hydrogen induced premature depletion advisory population.
 
Event or Problem Description
This product was returned to the manufacturer without any report of performance issues or adverse patient effects.The returned device was analyzed, and the product investigation result indicated that this device exhibited hydrogen induced premature battery depletion.See h10, for full investigation details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESSENTIO SR
Common Device Name
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112 5798
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19444346
Report Number2124215-2024-33314
Device Sequence Number20968827
Product Code LWP
UDI-Device Identifier00802526558900
UDI-Public00802526558900
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
N970003/S167
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Notification,Recall
Type of Report Initial
Report Date (Section B) 06/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date05/29/2017
Device Model NumberL100
Device Catalogue NumberL100
Device Lot Number703597
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2024
Initial Report FDA Received Date06/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/29/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
-
-