Model Number 170041
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/18/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant products: 170041, 170041 endo stitch tri 2-0 grn 18 s'dac, (lot#:j2a0645y), 170041, 170041 endo stitch tri 2-0 grn 18 s'dac, (lot#:j2a0645y), medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during laparoscopic diagnostic thoracoscopic image plicadura procedure, at the time of making a knot while repairing the diaphragm, the needle detached from the thread when the surgeon tried pulling the suture to check for breakage after finishing the cx.The issue occurred in three sutures from the same vial.The surgeon did not trust the knots and opted to start from scratch using a competitor's suture to resolve the issue.This led to 30 minutes or more extended surgical time.There was no patient injury.
 
Event or Problem Description
According to the reporter, during laparoscopic diagnostic thoracoscopic image plicadura procedure, at the time of making a knot while repairing the diaphragm, the needle detached from the thread when the surgeon tried pulling the suture to check for breakage after finishing the cx (surgery).The issue occurred in three sutures from the same vial.The surgeon did not trust the knots and opted to start from scratch using a competitor's suture to resolve the issue.This led to 30 minutes or more extended surgical time.There was no patient injury.
 
Additional Manufacturer Narrative
Correction: b5 additional information: g3 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted that the image: a segment of green suture could be observed in the center of the frame.Image: a segment of green suture and an endostitch needle were observed.The needle was disengaged from the suture.Image: two segments of green suture and an endostitch needle were observed.The needle was disengaged from the sutures.It was reported that the needle was detached.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGIDAC
Common Device Name
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19445176
Report Number9612501-2024-01347
Device Sequence Number8747586
Product Code OCW
UDI-Device Identifier20884521127002
UDI-Public20884521127002
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K934738
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number170041
Device Catalogue Number170041
Device Lot NumberJ2A0645Y
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/07/2024
Supplement Date Received by Manufacturer07/18/2024
09/03/2024
Initial Report FDA Received Date06/03/2024
Supplement Report FDA Received Date07/19/2024
09/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11