| Model Number |
PROXIMAL TENODESIS IMPLANT SYSTEM REV: 0 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023. arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The contribution of the device to the reported event could not be determined as the device is not expected to be returned for evaluation.The most likely cause for the reported failure can be attributed to misuse/mishandling due to damage to the device.As no further investigation was able to be performed no change in harm was identified.
|
| |
|
Event or Problem Description
|
|
It was reported that the button would not deploy.They used a peek implant to complete the case.
|
| |
|