• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON CLAREON VIVITY EXTENDED VISION UV ABSORBING IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON CLAREON VIVITY EXTENDED VISION UV ABSORBING IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS Back to Search Results
Model Number CCWET0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Impairment (2138); Visual Disturbances (2140); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Photophobia (4801)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A consumer reported that following the intraocular lens (iol) implant procedure, the patient experienced both near and distance vision.Patient could not see road signs and did not tolerate the fluorescent lighting.The patient also experienced glare and was unable to focus.The sight did not improve much and experienced mild posterior capsule opacification.The iol was centered.The patient was relying on glasses in order to sharpen her vision.There are two medical device reports associated with this file.This report is associated with the right eye.
 
Additional Manufacturer Narrative
The product was not returned.Product history records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause for the reported complaint.Each lens is subjected to a 100 percent assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.The reporter was the patient.The implanting surgeon suggested lasik or an exchanged to a monofocal.Information was provided that the patient received a second opinion and has opted to not proceed with lasik or with an iol exchange.Patient hopes to eventually adapt to the lenses.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information received that patient opted second opinion and doesn't want to proceed with any lasik or with iol exchange.She wants to keep lenses and eventually adapt to lenses.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON VIVITY EXTENDED VISION UV ABSORBING IOL
Common Device Name
EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key19446347
Report Number1119421-2024-01045
Device Sequence Number20959222
Product Code POE
UDI-Device Identifier00380652405782
UDI-Public(01)00380652405782(17)260730(11)220801(10)15463321(21)15463321007
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/30/2026
Device Model NumberCCWET0
Device Catalogue NumberCCWET0.205
Device Lot Number15463321
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/06/2024
Supplement Date Received by Manufacturer07/11/2024
Initial Report FDA Received Date06/03/2024
Supplement Report FDA Received Date08/08/2024
Date Device Manufactured08/01/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age71 YR
Patient SexFemale
-
-