| Model Number |
AU00T0 |
| Medical Device Problem Code |
Failure to Eject (4010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A health care professional reported that during an intraocular lens (iol) implant procedure, the lens did not load correctly in device and unable to implant.A new lens was opened and it had no issues.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3.,h.6 and h.11.The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.Viscoelastic was not provided.It is unknown if a qualified product was used.The product investigation could not identify a root cause.The account indicated the product was not available to evaluate.It is unknown if a qualified viscoelastic was used.The ifu (instructions for use) instructs: during device preparation and implantation of the company iol (intra ocular lens) with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Due diligence has been performed in an attempt to obtain further information on this event.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: 2024-35193.
|
| |
|
Search Alerts/Recalls
|