Model Number ZENITION 70
Medical Device Problem Code Radiation Output Failure (4027)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to philips that the device's wireless footswitch was inoperative.The device was in clinical use at the time of the event.No harm to the patient or user was reported.We are conservatively reporting this event as the investigation is ongoing.
 
Additional Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the system was not in clinical use at the time of the event.The philips field service engineer (fse) inspected the system onsite and confirmed that the wired footswitch was inoperative.To resolve the issue, fse replaced the wired footswitch, and the system was returned to good working condition.At the time this complaint was received, philips did not have enough information to exclude the potential for death or serious injury on recurrence, and as such the complaint was reported.Since that time, it was identified that the issue occurred without patient involvement and was identifiable prior to procedure commencement, which has led to the determination that this is scenario is not likely to cause or contribute to death or serious injury if it were to recur.As per the instructions for use, the user of the product must institute a user routine checks program.The user of the product shall make sure that all checks and actions have been satisfactorily completed before using the product for its intended purpose.It is instructed to not use the product for any application until the user is sure that the user routine checks have been satisfactorily completed, therefore, as the device was outside of use at the time the issue was identified, the user would identify this issue prior to use.Based on re-evaluation, this case is not reportable.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZENITION
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS INDIA LTD.
plot no. b-79, midc,
phase-ii, chakan
pune 41050 1
IN  410501
Manufacturer (Section G)
PHILIPS INDIA LTD.
plot no. b-79, midc,
phase-ii, chakan
pune 41050 1
IN   410501
Manufacturer Contact
derrick massuri
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key19446973
Report Number3003768277-2024-03128
Device Sequence Number3239896
Product Code OWB
UDI-Device Identifier00884838091528
UDI-Public00884838091528
Combination Product (Y/N)N
Initial Reporter CountryLO
PMA/510(K) Number
K183040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZENITION 70
Device Catalogue Number718133
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2024
Supplement Date Received by Manufacturer10/24/2024
Initial Report FDA Received Date06/03/2024
Supplement Report FDA Received Date11/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/12/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-