Model Number 9697
Medical Device Problem Code Missing Test Results (3267)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient/family (a1) was the initial reporter (e1), so personal information was not entered.No information was captured in sections a2 and a4 as the customer's age and weight were not provided.
 
Event or Problem Description
The customer called ascensia customer service to seek assistance with his contour next meter.During the troubleshooting, it was found that one of his blood glucose readings was not stored in the meter's memory.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
 
Additional Manufacturer Narrative
The customer returned the suspected contour next meter for evaluation.No test strips were returned.Therefore, an in-house testing was performed with the returned meter and in-house contour next test strips using blood sample, which gave a satisfactory performance.The blood glucose readings obtained with the in-house testing were properly stored in the meter's memory.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
Manufacturer (Section G)
PHC CORPORATION (REGISTRATION #: 3004167884)
110 oaza-inoshiri-aza
-nishiueno, wakimachi
mima tokusima 779-3 603
JA   779-3603
Manufacturer Contact
shweta gulati
100 summit lake drive
valhalla, NY 10595
9142361830
MDR Report Key19447952
Report Number1810909-2024-00084
Device Sequence Number8539914
Product Code NBW
UDI-Device Identifier00301939697010
UDI-Public00301939697010
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121190
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 05/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9697
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/09/2024
Supplement Date Received by Manufacturer06/26/2024
Initial Report FDA Received Date06/03/2024
Supplement Report FDA Received Date07/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-