| Model Number |
9390A-1001SP |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including functional stress testing without duplicating the report.An internal inspection found no discrepancies.The main board will be replaced as a precaution.The device will be recertified and returned to the customer once the repair estimate has been approved.Analysis of reports of this type has not identified an increase in trend.
|
| |
|