| Model Number |
DTMC1QQ |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) exhibited possible rapid battery depletion.The crt-d was explanted and replaced. no patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory showed the battery indicator signifying that it is time for device replacement.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported that the device battery depletion was believed to be normal and that replacement was prophylactic due to the device reaching elective replacement indicator (eri).
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.The returned device indicated shorting/low impedance in the battery.An alysis of the hybrid revealed the device set eri/rrt while implanted.Review of save to disk data showed an unexplained voltage drop beginning on 29 apr.2024 while the device was still implanted.Both loaded and unloaded pacing current drain levels were normal for this device over the range of battery voltage it operated under during its service time.There was no evidence of a high current drain condition on the hybrid and no hybrid anomalies were observed.Analysis of the battery revealed plated lithium was found to initiate from the battery case, extend under the head space insulator, and contact the cathode tab.Based on this, the premature battery depletion was determined to be the result of a plated lithium short between the battery case (negative polarity) and the cathode tab (positive polarity).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|