(b)(4).This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 analysis summary: the product was returned to ethicon for evaluation.The returned product determined that it was received, one labeled winding former with a partially dispensed suture and a detached needle that pertained to product code vcpb946h.Upon visual inspection of the returned sample, no breakage suture was observed.The sample was tested to verify the dispensability of the suture and it was not possible due to the suture being found hard to release from the tray by applying excessive force.This condition caused the needle to be detached from the suture.This product code contains an absorbable suture.Since the sample was received open, the functional test could not be performed due to undetermined exposure time to the environment.Based on the information currently available, the indispensable suture was identified during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.A review of the batch manufacturing records was conducted, and no related non-conformances were identified.
|