ABBOTT IRELAND DIAGNOSTICS DIVISION ALINITY I CEA REAGENT KIT; SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN Back to Search Results
Catalog Number 07P62-20
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/04/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed discrepant alinity i cea results generated for a patient sample.The following data was provided: patient 1 (b)(60 2023 result with old lot 49146fn00 = 2.0, (b)(6) 2023 result with new lot 50043fn00 = 3.1 additionally, the customer also noted a shift up in quality controls.No impact to patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A final report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
 
Additional Manufacturer Narrative
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed as returns were not available.Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.The ticket search by lot indicates that the complaint lot performs as expected for this product.Ticket trending review did not identify any trends.Device history review did not identify any non-conformances or deviations with the complaint lot.Labeling was reviewed and sufficiently addresses the customer¿s issue.Based on the investigation, no systemic issue or deficiency of the alinity i cea reagent lot 50043fn00 was identified.
 
Event or Problem Description
The customer observed discrepant alinity i cea results generated for a patient sample.The following data was provided: patient 1: on (b)(6) 2023: result with old lot 49146fn00 = 2.0, on (b)(6) 2023 result with new lot 50043fn00 = 3.1.Additionally, the customer also noted a shift up in quality controls.No impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY I CEA REAGENT KIT
Common Device Name
SYSTEM, TEST, CARCINOEMBRYONIC ANTIGEN
Manufacturer (Section D)
ABBOTT IRELAND DIAGNOSTICS DIVISION
finisklin business park
sligo F91VY 44
EI  F91VY44
Manufacturer (Section G)
ABBOTT IRELAND DIAGNOSTICS DIVISION
finisklin business park
sligo F91VY 44
EI   F91VY44
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key19454372
Report Number3008344661-2024-00065
Device Sequence Number8545714
Product Code DHX
UDI-Device Identifier00380740130855
UDI-Public(01)00380740130855(17)240813(10)50043FN00
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K990774
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/13/2024
Device Catalogue Number07P62-20
Device Lot Number50043FN00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/15/2024
Supplement Date Received by Manufacturer07/31/2024
Initial Report FDA Received Date06/03/2024
Supplement Report FDA Received Date08/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/12/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY I CEA REAGENT KIT, 07P62-20, (B)(6).; ALNTY I CEA REAGENT KIT, 07P62-20, (B)(6).; ALNTY I PROCESSING MODU, 03R65-01, (B)(6).; ALNTY I PROCESSING MODU, 03R65-01, (B)(6).
Patient SexUnknown
-
-