| Model Number |
60005 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
05/08/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device history record for the manufacturing lot was reviewed and there were no discrepancies or unusual findings.Manufacturer reference #: (b)(4).
|
| |
|
Event or Problem Description
|
|
A bilaterally implanted patient's light adjustable lens (lal, sn (b)(6), +26.0d) was implanted in the patient's right eye (od) on (b)(6) 2023.The patient completed one light adjustment treatment during the postoperative period.The patient did not complete the required lock-in light treatments.On (b)(6) 2024, approximately 10 months after the implant surgery, the lal was explanted from the patient's right eye.A new light adjustable lens was implanted as a replacement.Rxsight's first awareness of this event was on 05/08/2024.Reference report #3012712027-2024-00092 for the report on the left eye (os).
|
| |
|
Additional Manufacturer Narrative
|
|
The lal was cut into fragments during explantation.Visual inspection on the returned lens fragments found no device related issues.Due to the condition of the lens, further evaluation could not be performed.Section h6 codes were updated accordingly.A correction was made to the investigation findings code, the code 4248 was removed.Manufacturer reference #: (b)(4).
|
| |
|