| Catalog Number |
ROB10015 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during a cori assisted tka surgery, the silicone packing, which prevent substances entering into the ri robotic drill attachment, has been removed from the device.The surgery was completed, after a non-significant delay, with a s+n back-up device.No injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the ri robotic drill attachment, part number rob10015, serial number: (b)(6), used for treatment was not returned for evaluation, therefore a device analysis was unable to be performed.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.A historical escalation event review was not completed.The product was not returned and no evidence was made available to link the complaint to an escalation event.We have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with drill attachment wear and tear.Based on the investigation, no containment or corrective action is recommended or required at this time.Should any additional information be received, the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
|
| |
|
Search Alerts/Recalls
|