This device is classified as import for export, therefore 510k is not applicable.Model fb-15v is available in the usa with a 510k number k951199.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the biopsy inlet t-piece broken accessory blocking inside biopsy t-piece.Based on the result, we concluded that it was caused due the inadequate/insufficient reprocessing at the facility on the biopsy inlet t-piece.In addition, our technician confirmed that the insertion flexible tube buckled, and the insertion flexible tube crushed; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0588(channel)"" and/or the risk analysis results, it was evaluated to submit mdr.
|