| Model Number |
5386 |
| Medical Device Problem Code |
Unexpected Mode Switch (4076)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented remotely via merlin.Net.Upon review of the transmission, it was noted that the pacemaker was found in backup mode.No intervention was performed.The patient condition was unknown.
|
| |
|
Event or Problem Description
|
|
New information received notes the pacemaker was explanted.The patient condition was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of backup operation could not be confirmed.The device had no output, and no telemetry upon receipt.Device image could not be saved.Visual inspection after cutting open the device found no anomaly.Electrical analysis performed after attaching a new battery indicated normal functionality.Further evaluation of the output signal detected normal pacing parameters.Longevity assessment was performed, based on nominal conditions, and the device exceeded projected longevity indicating normal battery depletion.
|
| |
|