G2: the country of the event is india.H3.Device evualuation summary: product analysis #(b)(4).:part # g6626526 lot # h5823677 visual and optical inspection revealed the implant was returned damaged.The implant has cracked on both sides next to where the release/lock mechanism is.The implant is fragile and can be overloaded.It appears the implants structural integrity was overloaded during the implantation procedure.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|