| Model Number |
FIBERTAK |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The device was not returned for evaluation and therefore the alleged complaint event could not be confirmed.The most likely cause for the reported failure can be attributed to user-applied excessive force, improper bone prep; misaligned insertion; prying/leveraging the device during insertion.
|
| |
|
Event or Problem Description
|
|
It was reported that during a mpfl surgery the implants were pulled out of the drill hole.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a different device.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|