EXACTECH, INC. LOGIC CR TIB INSERT STD, SZ 5, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-47-5009 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Loss of Range of Motion (2032); Osteolysis (2377); Ankylosis (4527); Implant Pain (4561); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Joint effusion (4849)
|
| Date of Event |
05/02/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that approximately 102 months after a right total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, pain, swelling, decreased range of motion, effusion, osteolysis and bone loss.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
2857160 02-010-03-0350 - logic cr femoral cem, right, sz 5 2951620 02-012-45-5050 - lgc tibial fit tray cem sz 5f / 5t 3808539 200-02-35 - three peg patella 35mm the product associated with the reported event is within the scope of recall z-0021-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: corrected the following: health effect - clinical code, medical device problem code, type of investigation.Manufacturer narrative updated: the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, decreased range of motion, or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.
|
| |
|