It was reported that, during reprocessing, the videoscope tested positive for 1 colony forming units (cfus) of microorganisms that can be revived at 30° c and bacillaceae was detected.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
|
Updated: d9, h3, h4, h6, h11 corrected: b3 this report is being supplemented to provide additional information based on results of third-party testing, the customer provided cleaning disinfection and sanitization (cds) practices, the device evaluation, and the legal manufacturer's final investigation.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2024; sampling from: all channels; cfu: <1cfu/endoscope, conform; bacterial identification: no detection.Upon review of the information provided by users regarding the facility cleaning disinfection and sanitization (cds) practices, no obvious deviation from ifu was confirmed.The device was evaluated and no abnormalities were identified that could have led to the positive culture.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing, however, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: an insufficiently reprocessed endoscope and or accessory may pose an infection control risk to the patients and/or operators who touch them.Olympus will continue to monitor field performance for this device.
|