• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-N180
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The olympus scope was sent to an independent laboratory for culture testing and results are still pending.The investigation is ongoing and follow up with the customer is currently being performed.After culture testing, the device will be evaluated.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event or Problem Description
It was reported that, during reprocessing, the videoscope tested positive for 1 colony forming units (cfus) of microorganisms that can be revived at 30° c and bacillaceae was detected.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Additional Manufacturer Narrative
Updated: d9, h3, h4, h6, h11 corrected: b3 this report is being supplemented to provide additional information based on results of third-party testing, the customer provided cleaning disinfection and sanitization (cds) practices, the device evaluation, and the legal manufacturer's final investigation.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2024; sampling from: all channels; cfu: <1cfu/endoscope, conform; bacterial identification: no detection.Upon review of the information provided by users regarding the facility cleaning disinfection and sanitization (cds) practices, no obvious deviation from ifu was confirmed.The device was evaluated and no abnormalities were identified that could have led to the positive culture.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing, however, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: an insufficiently reprocessed endoscope and or accessory may pose an infection control risk to the patients and/or operators who touch them.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19461724
Report Number9610595-2024-11231
Device Sequence Number12446240
Product Code FDS
UDI-Device Identifier04953170307508
UDI-Public04953170307508
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K100584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGIF-N180
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2024
Supplement Date Received by Manufacturer07/16/2024
Initial Report FDA Received Date06/04/2024
Supplement Report FDA Received Date07/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/29/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-