| Catalog Number |
050-95018 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak during pd treatment.The patient reported noticing fluid dripping at the adapter set connection during fill four of treatment.It is unknown at which point in therapy the leak may have begun.Fluid was not discovered on the cassette inside the cassette door and there was no fluid in the pump module area of the cycler.Upon follow up, the patient reported that there were not any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient was able to complete treatment that night.The adapter is not available for return for physical evaluation by the manufacturer.The patient is using the same cycler with no further issues.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak during pd treatment.The patient reported noticing fluid dripping at the adapter set connection during fill four of treatment.It is unknown at which point in therapy the leak may have begun.Fluid was not discovered on the cassette inside the cassette door and there was no fluid in the pump module area of the cycler.Upon follow up, the patient reported that there were not any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient was able to complete treatment that night.The adapter is not available for return for physical evaluation by the manufacturer.The patient is using the same cycler with no further issues.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty cycler sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|