• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device evaluation is ongoing.Prior to the device evaluation, the device was sent out for additional microbiological testing.The hygiene microbiological investigation report indicated the channels of the scope were cultured and found less than 1 colony forming unit (cfu) of an unspecified revivable micro-organism.The results obtained comply with the target level for an endoscope subjected to high level disinfection and rinsed with sterile water.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Event or Problem Description
It was reported that during reprocessing, the cystonephrofiberscope tested positive for 1 colony forming unit (cfu)/120 ml of an unspecified revivable micro-organism.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacture's (lm) review of the customer cleaning disinfection and sterilization (cds) processes, the device evaluation and the lms final investigation.The lm reviewed the customer provided the cds processes where no obvious deviations from instructions for use (ifu) were identified.The device was evaluated where no abnormalities were found that could have led to the reported event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the ifu before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: ¿cyf-5 has an instruction manual for cleaning, and reprocessing method is described in the instruction manual for cleaning.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Common Device Name
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19463463
Report Number3002808148-2024-05061
Device Sequence Number20496367
Product Code FAJ
UDI-Device Identifier04953170292323
UDI-Public04953170292323
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K221690
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2024
Supplement Date Received by Manufacturer06/27/2024
Initial Report FDA Received Date06/04/2024
Supplement Report FDA Received Date07/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-