Additional information: d9, g3, h3, h6 complaint allegation is not confirmed.Upon visual inspection, the fiberwire® #2 braided polyblend suture, blue, w/ tapered needle returned did not have any issues with the sutures and/or needle attached.The braid component and needle pull test passed the requirements per the certification received from the supplier and arthrex records.However, the broken sutures/needles were not returned, nor was a photograph provided for a proper investigation.No issues were found with the packaging materials.The most likely cause of the reported failure can be attributed to user error of the device due to excessive force during suture passing, tightening, or tensioning, as well as incorrect bone preparation or contact with sharp instruments during use.
|