EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 2 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-35-2009
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Synovitis (2094); Joint Laxity (4526); Implant Pain (4561); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/17/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitants: 10 0025 13 016 a10007 - gps knee implant kit.2063701 200-02-29 - three peg patella 29mm.2135615 02-012-45-2020 - lgc tibial fit tray cem sz 2f / 2t.2435625 02-010-01-0220 - logic femoral ps cem left sz 2.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that approximately 111 months after a left knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear.It was observed that the polyethylene tibial insert was abnormally worn with delamination of the polyethylene and the patient experienced pain, swelling, and instability.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and instability or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.H6: corrected health effect and medical device clinical codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA PS MOD INSRT SZ 2 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key19464426
Report Number1038671-2024-01738
Device Sequence Number15316495
Product Code JWH
UDI-Device Identifier10885862173478
UDI-Public10885862173478
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2020
Device Catalogue Number02-012-35-2009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2023
Supplement Date Received by Manufacturer01/27/2025
Initial Report FDA Received Date06/04/2024
Supplement Report FDA Received Date01/27/2025
Date Device Manufactured05/17/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-