| Catalog Number |
ARS655200 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/08/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided on a supplemental report.
|
| |
|
Event or Problem Description
|
|
A patient experienced a dislocation due to a loose proximal body of the revive stem, potentially caused by a backed-out assembly screw and locking cap.Revision surgery was performed, replacing all components except for the distal stem.Additional implants used included two spacers, a low offset humeral tray, and a poly insert.
|
| |
|
Search Alerts/Recalls
|