• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER ORA SYSTEM WITH VERIFEYE MESSAGING; ABERROMETER, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER ORA SYSTEM WITH VERIFEYE MESSAGING; ABERROMETER, OPHTHALMIC Back to Search Results
Catalog Number 8065998329
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/30/2024
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported that an inaccurate measurement resulted in the incorrect intraocular lens being implanted at the wrong axis in the patient's right eye, after intraocular lens implantation surgery.
 
Additional Manufacturer Narrative
H.3., h.6.: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A manufacturing device history record review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.Service history was reviewed for the system.A service record relevant to the reported complaint was found.Service record was opened to address the reported event.An alcon representative performed an on-site assessment and was unable to confirm or replicate the reported event.The system was then tested per the service test procedure and found to meet company specifications.Based on the information obtained, the root cause of the reported event is inconclusive.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORA SYSTEM WITH VERIFEYE MESSAGING
Common Device Name
ABERROMETER, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key19467390
Report Number2028159-2024-00827
Device Sequence Number7498595
Product Code NCF
UDI-Device Identifier00380659983290
UDI-Public(01)00380659983290(17)310227(11)210301(10)13TKKV
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number8065998329
Device Lot Number13TKKV
Was Device Available for Evaluation? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/15/2024
Supplement Date Received by Manufacturer08/14/2024
Initial Report FDA Received Date06/05/2024
Supplement Report FDA Received Date09/10/2024
Date Device Manufactured03/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ORA SYSTEM WITH VERIFEYE MESSAGING, CART
Patient Age74 YR
Patient SexFemale
-
-