• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA THE PERSONALIZED KNEE SYSTEM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA THE PERSONALIZED KNEE SYSTEM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED Back to Search Results
Catalog Number 42556807005
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Osteopenia/ Osteoporosis (2651)
Date of Event 05/14/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: item #: 42542007901, lot #: 64960930.Item #: 42504607001, lot #: 64620016.Item #: 42560007514, lot #: 65028876.Item #: 42556607010, lot #: 64628788.Item #: 42556607010, lot #: 64783312.Item #: 42545000508, lot #: 64629414.Item #: 42560007514, lot #: 65028876.Item #: 42540200035, lot #: 65335632.Item #: 42512601014, lot #: 64403866.Item #: 66022663, lot #: 60041062.Item #: 66022663, lot #: 60041062.Item #: 66022663, lot #: 60041062.Item #: 66022663, lot #: 60041062.Item #: 66022663, lot #: 60041062.G2: foreign - event occurred in australia.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product was requested but not returned by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the patient underwent a revision procedure approximately one (1) year and nine (9) months post-implantation due to loosening of the implants and reported pain.The implants were found to be loose but well-adhered to the cement mantle.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d2; g1; g3; g6; h1; h2; h3; h4; h6; h10 no product was returned.Photograph provided shows explanted devices showing bony in growth on it.X-rays provided were reviewed and not submitted to mmi at this time as they are undated and it would be difficult to correlate to the onset of the alleged loosening and loosening can be difficult to assess from a single time point image.Review of device history records identified no related deviations or anomalies during manufacturing for all the reported devices.A definitive root cause cannot be determined.This complaint cannot be confirmed.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA THE PERSONALIZED KNEE SYSTEM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19468835
Report Number0001822565-2024-01888
Device Sequence Number16998798
Product Code JWH
UDI-Device Identifier00889024556942
UDI-Public(01)00889024556942(17)300916(10)64859452
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K191625
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number42556807005
Device Lot Number64859452
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2024
Supplement Date Received by Manufacturer11/11/2024
Initial Report FDA Received Date06/05/2024
Supplement Report FDA Received Date11/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NARRATIVE IN H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age75 YR
Patient SexMale
Patient Weight93 KG
-
-