SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. DELTEC TITANIUM PORT-A-CATH, LOW PROFILE; PORT AND CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Catalog Number |
21-4083-24 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/25/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
One device sample was received in used condition, in a plastic bag.During visual inspection of the device, it was detected that the catheter was broken since a catheter fraction remained on the port-catheter connection section, is also observed that the sleeve is locked/activated, this prevents catheter to detach from the port, if the catheter is being twisted or stretch while sleeve is locked, this could damage the catheter.The complaint was confirmed.The root cause cannot be associated with the manufacturing process since the failure could not be reproduced following assembly process, and per verification performed by production and quality personnel.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.No corrective actions were taken since the root cause could not be attributed to the manufacturing process.
|
| |
|
Event or Problem Description
|
|
It was reported that an interventionalist performs an infusion port implantation procedure on the patient, and the patient underwent chemotherapy for nearly a year after surgery.Recent findings of patient discomfort and poor infusion, and imaging revealed a broken catheter and subsequent intervention was performed to remove the catheter and remove the port.There was patient involvement and no patient harm/adverse event reported.
|
| |
|
Search Alerts/Recalls
|
|
|