| Model Number |
AVL ANGLED REAMER SLEEVE ASSEMBLY, 15 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint confirmed.Upon visual evaluation it was found that the angled reamer sleeve was received assembled to a reamer handle.The most likely cause for the reported failure can be attributed to user error of the device due to prying/leveraging the device during use.
|
| |
|
Event or Problem Description
|
|
On 08/05/2022, it was reported by a sales representative via sems that a ar-9676 driver shaft and ar-9297-15 reamer sleeve became stuck together.This occurred during an unspecified procedure with no patient harm.The case was completed using a backup tray.
|
| |
|